Bodystat Calibration Checker

Replacement Bodystat Calibration Checker to replace the one provided with your Bodystat device if lost or damaged.

Calibration checkers are common in industries where precise measurements are crucial, ensuring that instruments remain accurate over time. This device will verify the accuracy of your Bodystat body composition analyser, letting you know that your device is within the correct calibration range. 

Why measure Body Composition?

Measuring body composition is crucial for gaining a comprehensive understanding of an individual’s health and fitness status. Unlike traditional methods such as weighing yourself, which only provides overall body weight, body composition analysis offers a more nuanced insight into the distribution of different components within the body.

Who are Bodystat?

For over 30 years, Bodystat have been producing medical devices and software that utilise bioimpedance technology in the measurement of core cellular health to the critical acclaim of clinical physicians and international research institutions. Bodystat are dedicated to researching bioelectrical impedance technology to provide the best, most advanced electronics that accurately measure body composition and hydration within the human body, non-invasively.

Bodystat offer the latest advancements in medical science, an accuracy that is validated and a warm family service worldwide. Progress with purpose: For medical. For research. For fitness. For you.

Our relationship with Bodystat

Cranlea Human Performance have been working with the Bodystat team in the Isle of Man for over 20 years, helping delivering their vision to customers. Delivering Truly usable data that drives Performance, Fitness and Health.

Bodystat Calibration Checker

Calibration

Calibration of the Bodystat 1500MDD

The Bodystat 1500 should be calibrated at the beginning of a day of use

1. Attach one pair of red and black leads to any one terminal of the calibration unit.
2. Attach the other pair of red and black leads to the other terminal of the calibration unit.
3. Switch the unit on and enter any subject data or accept the “default” values on the display
4. Continue as normal to make the measurement.
5. Scroll using the down arrow to the values of impedance. You should find that the value reflects a reading of between 496 to 503, approximately a 0.5% variance on either side of the high precision 500 ohm resistor in the Bodystat calibrator.

Calibration of the Bodystat QuadScan 4000

The Bodystat QuadScan 4000 should be calibrated at the beginning of each day of use.

1. Attach one pair of red and black leads to any one terminal of the calibration unit.
2. Attach the other pair of red and black leads to the other terminal of the calibration unit.
3. Switch the unit on and make a note of the test number.
4. Accept the “default” values on the display.
5. Continue as normal to make the measurement.
6. Scroll using the down arrow to the values of impedance. You should find that the results at 5 kHz, 50 kHz, 100 kHz and 200 kHz should reflect readings of between 496 to 503, approximately a 0.5% variance on either side of the high precision 500 ohm resistor in the Bodystat calibrator.

Do you have any questions about Bodystat Calibration Checker?